激光多普勒血流灌注成像仪 PeriScan PIM 3


Type (protection against electric shock): Class I Equipment, Type B
Laser: EN 60825-1 Class 2
Wavelength: 670 nm
Max output power: 1 mW
Beam diameter: 1 mm
Mains: 100 - 240 VAC, 50 - 60 Hz
Current: 90 mA (240 VAC) and 220 mA (100 VAC)
Power consumption: 35 VA
Dimensions Scanner head:
Opto-isolation unit:
23 x 12 x 12 cm
24 x 20 x 18 cm
Weights
Scanner head:
Opto-isolation unit:
Arm and stand:
 
2.5 kg
2,0 kg
7,4 kg
EC Declaration of conformity: MDD 93/42/EEC
Transport and storage conditions:  +10 to +50 °C at 10-90% RH
Operating conditions: +10 to +30 °C at 30-75% RH
Atmospheric pressure between 700 mbar and 1600 mbar
Computer requirements:
CPU 300 MHz or faster, 128 MB RAM
50 MB free hard disk space
Windows 98 or higher
Max scan area: Approximately 30 x 30 cm depending on distance to object
Scan depth: Approximately 0.5-1 mm depending on tissue properties
Scan times:
(Approximate)
(10 x 10 points) 9 seconds
(16 x 16 points) 18 seconds
(32 x 32 points) 1 minute
(64 x 64 points) 4 minutes